Specification Sheet

Model 105 Sigma-Delta Counts-to-Volts Converter

 

Other precision data
acquisition products



zoom
 
OVERVIEW
The Model 105 is a positive edge triggered counts-to-volts converter with a range of 0 to 380 KHz. The analog output is scaled to 1 volt at 200 KHz. Resolution can be as good as one count per half second when used with a sufficiently quiet high resolution A/D converter. The maximum voltage output is 2 volts. (380 KHz = 1.9 volts.)

The filter time constant for the Model 105 is 0.5 seconds. It settles to 18 bits in 4 seconds after an instantaneous full-scale step.

LINEARITY
Linearity is typically 0.001% of full scale.

POWER REQUIREMENT
+12 to 15 volts (regulated) at 15 ma and -12 to 15 volts (regulated) at 1 ma.

DIGITAL INPUTS
Input Connections (34-pin dual row header):
Pin 27 +15 VDC
Pin 29 -15 VDC
Pin 32 ground
Pin 33 input signal (5 volt logic)

DIGITAL OUTPUTS
Output Connections (screw driver terminals):
+ out to A/D + input
- out to A/D - input
ground use for common mode ground connection, if necessary (can also be used to terminate shield. Do not connect shield to ground at both ends.)
PRICE
$260

HOME

Contact us for sales or technical assistance   lawsnlab@lawsonlabs.com