Installation Instructions for L-EMF-DAC software.:
Under Win7 and above, it is imperative that the installation program, SETUP.EXE, is run as administrator. Simply having administrator
rights is not enough.
You must right click on the Setup file and select "run as administrator" from the menu.
If you are reinstalling after a failed installation, right click the UNINSTALL utility and choose "run as administrator", then repeat the installation, as above.
Once installed, you may also need to run the application as administrator, by right-clicking
the app and choosing the option to "run as administrator."
There are three grounding options selected by the mode switch above our logo:
DIFF is the best, for pure differential measurements. In the DIFF positions, you *must* add a
third connection to the solution. Use a separate wire to connect the solution to the COMMON terminal
or BNC connector.
For SAFE and COMMON positions, do not use the third wire. In these cases, the reference
input is connected internally to the common. The Safe mode protects the hardware against severe overvoltage and
may be just a bit noiser as a result. The SAFE and COMMON POSITIONS can have the
effect of lowering the input impredance on the reference side.
The input impedance is 1013 at room temperature. Input current is typically 3 picoamps. A 1015
version is available as an option. An 8 channel differential option is also available.
For more information about the internal function of the EMF16, refer to the Model 201 Manual. The
16 EMF channels are addressed at analog input channel number zero, external sub-channels 0 through 15.
NOTE: For demanding applications, you may want to consider the Model 201 graphics software which is
free at our website
HERE
back to top
|